Validation of biplane fluoroscopy system tracking using bone phantoms

This abstract has open access
Abstract Summary
Using our lab’s biplane fluoroscopy system, we collected data on eight bone phantoms to assess the accuracy of our custom bead-based and model-based tracking algorithms. Custom rigs were constructed to apply uniaxial rotation and translation to these bone phantoms. External potentiometer measurements provided a “gold standard” for the tracking algorithms. The bead tracking root-mean-squared errors are consistent with the results from our previous work (sub-mm and sub-deg accuracy). The rotational bead tracking for the remaining bone phantoms in addition to the model-based bone tracking is ongoing.