Novel device for lower limb weight-bearing dynamic ct: a 3d knee kinematic analysis

This abstract has open access
Abstract Summary
One of the main limitation of CT imaging and especially Dynamic CT acquisitions of the lower limb is the unavailability of devices that create weight-bearing (WB). This study aimed to validate a novel device developed for this proposes. 15 healthy subjects performed a series of orthostatic squats, followed by a series of “horizontal” squats on the device at different resistances based on body weight percentage (BW%). 3D-kineamtics and surface EMG were recorded. To explore the degree of similarity between both settings, the 3D-kinematics of left knees were compared using Linear Fit Method and mean differences. Very high average correlation was found for most BW%. Average rotational differences were below 10˚ for nearly all components. The novel device seems able to replicate knee kinematics when compared to a true orthostatic wall squat.